SW 6.16 INSTALLATION PROCEDURE
MT 8503 HANDLERS
- TAKE NOTE OF THE FOLLOWING SETTINGS:
- BIN CATEGORIES(CP88)
- HANDLER ID(CP44)
- POWER OF THE HANDLER
- REPLACE THE OLD CPU WITH THE URGRADED ONE.
- SET THE HANDLER TO CP 110.
- POWER ON HANDLER.
- SELECT HANDLER TYPE BY PRESSING START AND USING CP110
- MT8501HC/1 a - 8501 MODEL CAPABLE OF HI/LO TEMP SINGLE SITE
- MT8501H/1 b - 8501 MODEL CAPABLE OF HI TEMP SINGLE SITE
- MT8501A/1 c - 8501 AMBIENT HANDLER SINGLE SITE
- MT8503HC/1 d - 8503 MODEL CAPABLE OF HI/LO TEMP SINGLE SITE
- MT8503H/1 e - 8503 MODEL CAPABLE OF HI TEMP SINGLE SITE
- MT8503A/1 f - 8503 AMBIENT HANDLER SINGLE SITE
- MT8588HC/2 g - 8588 MODEL CAPABLE OF HI/LO TEMP DUAL SITE
- MT8508H/2 h - 8588 MODEL CAPABLE OF HI TEMP DUAL SITE
- MT8588A/2 i - 8588 AMBIENT HANDLER DUAL SITE
- MT8589HC/2 j - 8589 MODEL CAPABLE OF HI/LO TEMP DUAL SITE
- MT8589H/2 k - 8589 MODEL CAPABLE OF HI TEMP DUAL SITE
- MT8589A/2 l - 8589 AMBIENT HANDLER DUAL SITE
- MT8589HC/1 m - 8589 MODEL CAPABLE OF HI/LO TEMP SINGLE SITE
- MT8589H/1 n - 8589 MODEL CAPABLE OF HI TEMP SINGLE SITE
- MT8589A/1 o - 8589 AMBIENT HANDLER SINGLE SITE
- PRESS STOP TO SELECT THE HANDLER TYPE.
- HANDLER START BUTTON WILL STILL BE LIGHT. SO SET FUNCTION KNOB TO NORMAL THEN PRESS START. HANDLER WILL RECONFIGURE.
- AFTER INITIALIZATION, RETURN BACK SETTINGS OF BINNING(CP88) AND HANDLER ID (CP44) FROM STEP1.
- SET TEMPERATURE GUARDBAND TO 5 USING CP87 RB05.
NB. IF HANDLER DOES NOT ENCOUNTER ANY UNUSUAL PROBLEM, RELEASE HANDLER FOR PRODUCTION. ELSE, TAKE NOTE ON THE PROBLEM ENCOUNTERED, RETURN BACK THE OLD CPU AND LABEL THE NEW CPU WITH ITS CORRESPONDING PROBLEM
END/10/31/2000/JONG IGNACIO