MULTITEST Handler MT85XX Temperature Calibration Procedure Note: Handler must be empty Important: Make sure handler did not heat at all For 8503/88 kits only STEP 1: CP 22, RB 18(start 2X) Erase Compensation table for Input Magazine (Wait until cleared) STEP 2: CP 24, RB 03(start 2X) Erase Compensation table for Contact Site (Wait until transfer is done) For 8501 kits only STEP 2: CP 21(start), ST 10(start)- dedicate to this soaktime/memory for one kit IMPORTANT: SKIP STEP 8 for 8501 kits) STEP 3: Set Handler to 20 deg C if not to 30 deg C Record on Table(Use CP 70, 76 respectively for Input mag/Sensor 1 and Contact Site/Sensor 7) Do record on other Temperature whichever is applicable Note: For 20/30 deg C setting wait 15 minutes before recording For extreme temp wait for about 1 hour. STEP 4: Calculate Correction Value CORRECTION VALUE = EXTERNAL SENSOR RDG - DISPLAYED TEMP STEP 5: Enter correction values individually for Input Mag and Contact site per calibration points. (Push start to increase and stop to decrease) For input Mag: CP 22, RB 00 For Contact Site: CP 22, RB 06 Note: Do all correction value input for all specified calibration points. STEP 6: After completing all compensation value, change CP setting other than 22 to exit. STEP 7: CP 22, RB 19- Interpolate (Wait until int. pol. done) STEP 8: CP 24, RB 02 -transfer compensation values to CS1 contact site RAM (Wait until transfer is done)