      PERMEATE DIVERT

(NOTE: The RO system goes into Permeate Divert for two minutes when the RO first turns on. Sometimes the switch is delayed and will alarm briefly on the floor. Check to make sure the alarm does not go off on its own within 3-5 minutes first.)

          - Water conductivity is high.
              a. Check the conductivity on ECM-100 (Monitor on RO face) on bottom left of screen. Conductivity
                  should be below 10. If high, contact BEST.

          - Water quality is poor (>30).
              a. Check Final Water Quality Meter to make sure water quality number is below 30. If not, contact
                  BEST.