      TANK LOW

          - The Storage Tank has drained to a low level.
              a. If RO is running:
                  1. Check clinic to make sure there are no leaks in the loop or that water is not running somewhere
                  (check Reuse sinks).

                  2. Check RO room to make sure all normally closed valves are closed.
              b. If RO is not running.
                  1. Try pressing the Off pad and then pressing the On pad to restart the system.
                  2. Check ECM-100 (monitor on RO face) screen for alarm condition. If alarm condition exists refer to
                      other Troubleshooting information.

                  3. Call BEST.