PRE-TREATMENT:
CARBON TANKS


      * Total Chlorine is tested immediately after the first or primary carbon tank (or tanks). If the
         reading is less than 0.1 mg/L dialysis treatments may continue.

      * If the Total Chlorine level is greater than 0.1 mg/L the output of the secondary Carbon Tank
         should be tested immediately. If the reading is less than 0.1 mg/L dialysis treatments may
         continue and the following must be done:

            - Contact B.E.S.T. and your CD or RD.
            - Chlorine levels must be check EVERY 30 MINUTES and recorded on the Chlorine
               Breakthrough Log
.

      * If the Total Chlorine level is greater than 0.1 mg/L post the secondary Carbon Tank
         dialysis treatments must be stopped immediately and patients taken off the machines
         without returning their blood.

* CARBON TANKS MUST NEVER BE BYPASSED!!!! *